1 Sample mount ledge
Permits mounting of 1” and 2”
samples in %T and %R modes
08 101697 00
The Standard Sample Holder Kit will accommodate mounting of reflectance
standards, specifically:
Item Description
Protruding PTFE
reference plate
Large PTFE reference
plate
1 ¾” diameter
2” Reflectance standards
Reflectance standards (Spectralon) 2”, for use with Diffuse
reflectance accessories
1” color standards
Calibrated diffuse color standards set for DRA. Set includes
four 1” D diffuse color standards. Red, green, blue and
yellow
Click here for details of what is included in each of the other sample mounting kits.
Installation
You will need:
A flat blade screw-driver (is included in the instrument tool kit)
A 3 mm hexagonal ball driver
The Extended Sample Compartment
1. Turn off the spectrophotometer and remove any cell holders or other
accessories from the sample compartment. The PC can be left on.
2. Close the sample compartment, turn the spectrophotometer on and allow it to
initialize.
3. Open the accessory case and remove the metal clamp and alignment target.
4. Use the hex ball driver to remove the shipping hex screw from the the end cap
of the sphere.
5. Remove the accessory cable from the white cable clamp.
6. Remove the cover over the mirrors by loosening the two screws on top of the
cover and place to one side.