Mantis 5000 Especificaciones Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 49
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
Surface type Reflectivity Reference
Matt High PTFE reference plate
Matt Low
Labsphere diffuse
reflectance standards
Glossy High
Protruding PTFE
reference plate
Glossy Low NIST SRM 2021
The Troubleshooting section contains information on the available reference surfaces.
Figure 5: Highly diffuse samples are measured against a flat PTFE reference plate
Figure 6: The protruding PTFE reference plate more accurately emulates highly
specular samples
Inaccuracy in the standard
If the standard material used to calculate a value is not accurate, this will hamper the
determination of the reflectance of the sample. Great care must be taken to keep
reference materials clean and unscratched.
For a detailed discussion on the factors affecting the measurement of diffuse
reflectance and transmittance with integrating spheres, refer to reference (2).
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 48 49

Comentarios a estos manuales

Sin comentarios