14. Adjust the height adjustment screw on the side of the polarizer mount until the
transmission reading is half (+/- 1%) of the reading noted in step 12.
15. Open the sample compartment lid and rotate the aperture mask so that only the
upper aperture is open. Repeat the measurement until the reading for the upper
aperture is half (+/- 1%) of the reading when both apertures are open.
16. Use the Validate application to check the photometric accuracy of the DRA
with the double aperture attachment. Refer to the Validate
section of the help.
Operation
Samples can be mounted at the transmission or reflectance ports. At the reference
port, attenuators may be mounted. Paddles, clamps and a sample ledge are provided
you to assist with mounting your samples.
Figure 31. Front, the Paddle; rear, the
Clamp
Figure 32. The Ledge
The Varian internal DRA provides different mounting options:
standard sample holder kit;
liquid sample holder;
small sample holder;
powder cell;
small sample mount;
polarizer;
double aperture attachment
The following table illustrates which type of readings can be taken with which
attachments.
Attachment %T %R
Standard
Liquid sample holder
Small sample mount
Powder cell
Caution
When performing measurements and calibrations with the DRA some steps require