Figure 11: The Cary 4000, 5000 and 6000i Internal Diffuse Reflectance accessory
with the optics cover removed. The large reference plate is clamped over the
reflectance port at the centre of the sphere cap.
Warning
This accessory contains electrical circuits, devices, and components
operating at dangerous voltages. Contact with these circuits, devices and
components can cause death, serious injury or painful electrical shock.
To prevent electrical shock operators and other unauthorized personnel
must never remove the cover. This must be opened only by Varian-
trained, Varian-qualified or Varian-approved service engineers.
7. Position the accessory so that the two rear lock-down feet are situated at the
back of the sample compartment and gently lower the accessory so that the
front pin on the base plate fits into the post clamp on the sample compartment
floor. Ensure that the base of the accessory is parallel to the sample
compartment floor.
8. Slide the lever at the front of the sample compartment to the left, to lock the
base plate into position.
Caution
Do not plug the DRA into the spectrophotometer when the latter is turned on, or
electronic damage may occur.
9. Take the clamp post from under the foam in the wooden shipping case and
fasten it in the hole in the accessory cap. Place the sample clamp on the clamp
post.
Alignment
Warning
To avoid eye damage use care when looking into the various ports
during the alignment process. Do not look into the light beams.