Mantis 5000 Especificaciones Pagina 36

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 49
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 35
1. Click the Windows Start button, then Programs then Cary WinUV and then
Scan.
2. Click the Setup button.
3. Set the following parameters:
Cary tab
X mode: Mode
N
anometers
X mode: Start 1800
X mode: Stop 300
Y mode: Mode % R
Y mode: Y min -5.00
Y mode: Y max 110.00
Show status display ON
Options tab
SBW/Energy: Beam mode Double
SBW/Energy: Slit height Reduced
Source/Detector: Lamps UV-Vis
Source: Source changeover (nm) 350.0
Source: Detector changeover (nm)800.0
Independent tab
Independent control ON
Measurement mode Auto
UV-Vis controls
Ave time (s) 0.1
Data interval (nm) 1.000
Scan rate (nm/min) 600.00
SBW (nm) 2.00
N
IR controls
Ave time (s) 1.000
Data interval (nm) 2.000
Scan rate (nm/min) 120.00
Energy level 3.00
Baseline tab
Correction Zero correction
Autostore tab
File Storage: Storage Storage on (prompt at start)
Note
The NIR detector is noisier than the UV-Vis detector, so a higher Signal
Averaging Time (Ave time) may need to be set in the NIR. To obtain a constant
signal-to-noise ratio over the entire wavelength range, set a value in the Signal-
to-noise field. If you require a faster scan rate, increase the Data interval or
lower the Ave Time.
Vista de pagina 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 48 49

Comentarios a estos manuales

Sin comentarios