Figure 25: DRA socket
21. Click the Windows Start button, then Programs, then Cary WinUV, and then
Validate. Turn on the spectrophotometer. If the instrument will not initialize, proceed to
the Troubleshooting
section.
22. Close the Validate application.
23. If desired, confirm the proper operation of the accessory by performing the
diagnostic scans described in the Diagnostic Scans
section of the online help. You may
wish to perform these tests on the initial installation of the DRA.
24. Run the Calibration
checks.
Warning
This accessory contains electrical circuits, devices, and components
operating at dangerous voltages. Contact with these circuits, devices and
components can cause death, serious injury, or painful electrical shock.
To prevent electric shocks, operators and other unauthorized personnel
must never remove the main cover. This must be opened only by Varian-
trained, Varian-qualified, or Varian-approved service engineers.
Alignment
Warning
To prevent eye damage use care when looking into the various ports during
the alignment process. Do not look into the light beams.
After installing the DRA it is necessary to check the alignment of the accessory. In some
cases it will then be necessary to Align the DRA
.
You will need:
• A flat blade screw-driver
• 3mm Allen key (provided)