Troubleshooting
Warning
This instrument contains electrical circuits, devices, and components
operating at dangerous voltages. Contact with these circuits, devices and
components can cause death, serious injury, or painful electrical shock.
To avoid electrical shock operators and other unauthorized personnel must
never remove the main cover. This must be opened only by Varian-trained,
Varian-qualified, or Varian-approved service engineers.
Problem Solution
Measurements on
standard materials are
not producing certified
values.
Ensure that you are referring to the hemispherical reflectance
values of the SRM, not the directional and hemispherical values.
Ensure that your PTFE disks are clean and in good condition.
If you are calibrating the PTFE disk, you need only apply the
necessary correction to your measured values.
The signal-to-noise
ratio is low.
This can be caused by degraded sphere coating, poor detector
performance, the lamps requiring replacement, unclean mirrors
or lens, or bad alignment.
Check the sample port image, and re-align the accessory if
necessary.
Check that the lens and mirrors are clean, and carry out a Zero
%R error check.
A Failure to initialize
instrument error
message occurs.
Error number:
9:9303SEQ:E9303
Check that the:
The DRA is plugged in.
The beams are not blocked.
The DRA is not out of alignment.
A Cary Exception error
occurs.
An ADL command is necessary to clear this error. Click here to
view/print a PDF instructing you how to use this ADL.