Mantis 5000 Especificaciones Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 88
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
Alignment
Upon first time use, the sample holder should be loaded into the centre-mount port and
checked for proper angle alignment to the sample beam, the beam should strike at normal
incidence when the dial is set at 0°. If it does not, rotate the knob until the angle is normal
to the sample beam, loosen the four screws along the perimeter of the dial and rotate the
dial so it reads 0°.
The reflecting surface of the standard should be the same angle of incidence as the
intended sample. When properly aligned, the 0° dial reading on both clip style and jaw
style sample holders holds the sample for normal incidence where the reflecting surface
is perpendicular to the path of the sample beam.
Ideally, the beam geometry for a reflectance accessory should be optimized for the
application concerned. For centre-mount applications, the ideal configuration has the
sample beam focused on the front surface of the sample at the centre of the sphere. In
general, the maximum dimensions of a centre-mounted sample is proportional to the size
of the integrating sphere. It should be obvious that the smaller the sample, the more
accurate the reflectance measurements - as long as the sample beam does not overfill the
surface area of the sample. For measurements with an Edwards sphere, ASTM E 903
recommends limiting specimen size to one percent of the surface area of the sphere.
Thus, for a 150 mm diameter integrating sphere, the maximum specimen size is a total
surface area of approximately 700 mm
2
. Sample sizes exceeding the recommended
specimen dimensions may interfere with the radiance uniformity at the wall surface or
clip the reference beam inside the integrating sphere.
Centre-mount sample holders inflict further restrictions on reflectance samples. The DRA
accessories utilize a three position mirror to focus the sample beam at the transmission
port, sample reflectance port, or centre-mount positions, depending on the sphere
configuration. Mirror M3, however, is a concave mirror such that the spot produced by
the sample beam depends on the position of the mirror and the angle of incidence to the
sample. The spot size grows larger as the angle of incidence increases, to a point where
the spot may no longer fit on the sample surface. At near-normal angles, the specular
component of the reflected beam is rejected completely out the transmission port and is
not collected during the scan. Finally, the horizontal dimension of the sample or
reflectance standard may clip the reference beam at steep angles of incidence, thereby
placing premature limitations on reflectance measurements.
Loading a sample into the jaw style centre sample holder
Tighten the jaws using the thumbscrew. Do not overtighten the jaw mechanism as this
could damage the sample.
When loading a sample into the clip sample holder, the sample should be held perfectly
vertical by the clip with the plane of the reflecting surface along the centre of rotation of
the dial. The sample beam path will strike the sample at a point midway between the
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 87 88

Comentarios a estos manuales

Sin comentarios