Data interval (nm) 1.000
Scan rate (nm/min) 600.00
SBW (nm) 2.00
NIR controls
Ave time (s) 1.000
Data interval (nm) 2.000
Scan rate (nm/min) 120.00
Energy level 3.00
Baseline tab
Correction Zero correction
Autostore tab
File Storage: Storage
Storage on
(
p
rom
p
t at start)
Note
The NIR detector is noisier than the UV-Vis detector, so a higher Signal Averaging
Time (Ave time) may need to be set in the NIR. To obtain a constant signal-to-noise
ratio over the entire wavelength range, set a value in the Signal-to-noise field. If
you require a faster scan rate, increase the Data interval or lower the Ave Time.
4. Place the appropriate PTFE reference plate over the sample port.
4. Place the appropriate PTFE reference plate over the sample port.
5. Press the ’Goto’ button to open the ’Goto wavelength’ dialog. Enter ’801’ in the
wavelength field and click OK. The instrument will then drive to 801 nm.
Note
Note the SBW reading in the Status Display dialog.
6. The SBW reading needs to be at approximately 18. This can be adjusted by changing
the NIR Energy level in the Setup page, under the Independent tab. If the reading is
greater than 18, raise the value of the Energy level field. If the SBW is under 18, lower
the energy level. Once the Energy level has been adjusted, you will need select OK in
order to note the change to the SBW reading. Then press the Goto button and again drive
the instrument to 801 nm by entering ’801’ in the wavelength field and pressing OK. This
procedure may need to be repeated in order to reach the required setting. Please note that
each time you adjust the Energy level reading, you must also drive the instrument to
801 nm.