Mantis 5000 Especificaciones Pagina 84

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 88
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 83
1. Remove the accessory from the wood container and set it on a flat surface.
2. Check that each mirror mount in the optics chamber is fastened securely to the base
plate.
3. Blow clean air or nitrogen over the transfer optics to remove any dust lying on the
mirror surfaces.
4. Examine the surface of each mirror using a flashlight. Look for dust particles or film
damage that might scatter the incident radiation.
5. Remove the reflectance standards and centre mount plug from the integrating sphere
or box. Inspect the reflective surfaces of the standards for damage or dirt. Use a
magnifying glass, if necessary.
6. Inspect the reflecting surface of the centre mount plug.
7. Illuminate the sphere interior with a flashlight and inspect the surfaces for soiling.
Blow out any debris using clean dry air or nitrogen.
8. Re-install the centre mount plug and reflectance standards to their respective sphere
ports or store the entire accessory in the wooden container.
Mirror Cleaning Procedure
Hot Tip
Holding the mirror under a bright light will show up any marks on the surface of
the mirror.
The transfer optics mirrors should not be touched or handled with bare fingers. Never
clean the mirrors with abrasive cleaners or tissue paper. The mirrors have a protective
magnesium fluoride coating to allow cleaning if necessary. The following cleaning
procedure is recommended:
1. Remove the mirror fixture.
2. Spray the mirror with Fantastic Spray Cleaner.
3. Rinse the mirror with distilled water.
4. Repeat until the water sheets off the mirror.
5. Blow dry the mirror surface with a stream of clean nitrogen.
6. Reinstall the mirror fixture and check the alignment of the optics.