Mantis 5000 Especificaciones Pagina 73

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 88
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 72
Applying a Nitrogen Purge
The Cary 4000/5000/6000i instruments are fitted with connection points for purging the
optical system with nitrogen to enhance the performance of each instrument at extremes
of its range. More details are provided in the Cary Hardware operation manual (part
number 8510197200), supplied with the instrument.
Nitrogen supplies are not available from Varian but may be obtained from commercial
suppliers. Liquid nitrogen (in conjunction with a heat exchanger) is recommended
because it is generally less costly than compressed nitrogen and is of better quality.
Where compressed nitrogen must be used, the gas must be dry, oil–free and
uncontaminated. Do not use compressed nitrogen from a supplier who uses oil or water in
the compression process (these methods leave fine particles of oil or water suspended in
the nitrogen that may be deposited on the instrument optics). Only use nitrogen from a
supplier who fills containers from immersion pumps lubricated with liquid nitrogen.
Note
The instrument warranty will be void if damage is caused by the use of contaminated nitrogen.
Operating pressure at the pressure regulator (# 7 in figure 53) for the nitrogen purging
system is 83 to 172 kPa (12 to 25 psi). Use a suitable regulator and gauge assembly to
ensure that the nitrogen supply is maintained at the correct pressure.
Nitrogen supply tubing should be clean, flexible plastic tubing 6 mm (1/4“) inside
diameter (Tygon PVC or equivalent). Do not use rubber tubing as this is usually treated
internally with talc which will be carried into and contaminate the instrument optics.
The nitrogen system should include a manifold assembly with inlet from the supply and
two outlets for connection to the instrument. Manifold outlets should each be fitted with a
stop valve and flowmeter for control of gas flow to the instrument. Flow meters should be
adjustable for flow rates of 0 to 30 litres per minute (0 to 64 cubic feet per hour). Refer to
the figure below for more details.
Figure 53. The position of flowmeters when purging with nitrogen:
Vista de pagina 72
1 2 ... 68 69 70 71 72 73 74 75 76 77 78 ... 87 88

Comentarios a estos manuales

Sin comentarios