Figure 43. The double aperture attachment on the polarizer mount.
The accessory optics should be aligned according to the Installation instructions before
installing the double aperture. The base mount for the double aperture apparatus is the
same base used for the polarizer
. Install the double aperture accessory as follows:
1. Remove the DRA from the instrument sample compartment. Locate mirror M3 at the
"R" position.
2. Insert the base mount into the locator holes in the DRA base plate as shown in the
figure above. The orientation of the mount is not important. Fasten the device to the
accessory using the M5 socket head screw provided and a 4 mm hex wrench.
3. Re-install the DRA into the sample compartment.
4. Click the Windows Start button, then Programs, then Cary WinUV, and then Align.
5. Turn on the instrument and allow the instrument time (about 2 hours) to warm up.
6. Remove the reflectance standards at each port.
7. Open the sample compartment lid and slide the aperture plate into the base mount slot
as shown in the figure. The protruding alignment pins should point into the accessory.