7. Flick the lockdown lever on the front instrument panel to the left.
8. Install the guillotine/light seal.
9. Load a reflectance standard at the reference and sample reflectance ports. Remove any
centre-mount sample holders installed in the sphere and install the centre port plug.
10. Connect the accessory detector cable to the connector at the rear of the sample
compartment. Make sure the detector cable does not obstruct the beam path inside the
optical chamber of the accessory. Install the accessory cover.
11. Click the Windows Start button, then Programs, then Cary WinUV, and then Scan.
12. Close the sample compartment lid and turn power to the instrument ON.
13. Allow the instrument and accessory to warm up for at least 30 minutes before
proceeding with measurements.
Removing the DRA from the instrument
Use the following procedure to disconnect and remove the DRA from the instrument:
1. Turn the spectrophotometer off.
2. Open the sample compartment lid and disconnect the accessory cable.
3. Remove the accessory cover.
Figure 46. Removing the accessory cover.
4. Remove the guillotine/light seal.
Vista de pagina 54