3. Click the Setup button and check the parameters. Normally, the setup parameters for
the application should match those used for the baseline scan. If using a previously
defined baseline correction, retrieve the correct baseline.
4. Click the Start button.
5. If the standard reference option was used for the baseline scan, the reflectance spectra
is generated automatically. If the simple or zero baseline correction was selected,
multiply the sample scan data by the spectral reflectance factors for the calibrated
reflectance standard.
0°/h Transmission Measurements (Comparison Method)
For transmission measurements, make sure that mirror M3 is placed in the transmission
position, marked "T" on the base plate, located farthest from the sphere. If this mirror is
moved or if the cuvette holder is installed over the transmission port, the transfer optics
must be realigned. Repeat the alignment procedure if M3 has been moved. The "T"
position mirror configuration provides a condensed spot size of approximately 5 mm
wide x 9.5 mm high. If a large transmission spot size is desired, such as for the
measurement of fabrics or other homogeneous samples, place mirror M3 in the
reflectance "R" position. This mirror configuration provides a spot size of approximately
11 mm wide x 28 mm high. If an intermediate spot size is required, place the mirror in
the centre-mount "C" position.
1. Collect a Baseline scan
, following the procedures specific to transmission
measurements. If the sample resides on a substrate, load an untreated substrate blank at
the transmission port for the baseline scan recordings. If the transmission sample is
suspended in solution, load a blank cuvette at the transmission port for the baseline.
2. Load the sample at the transmission port. Do not change the configuration of the
integrating sphere in any other way.
3. Click the Setup button and check the parameters. Normally, the setup parameters for
the application should match those used for the baseline scan. If using a previously
defined baseline correction, retrieve the correct baseline.
4. Click the Start button.
5. The value displayed by the instrument is the sample transmission. No data correction
is required.
0°/d Transmission Measurements (Comparison Method)
The approximate diffuse transmission of a given test sample can be measured using the
DRA as follows: