Figure 27. Checking the beam at the reference port.
5. Mount or hold the translucent paper directly in front of the transmission port. Check
that the sample beam is confined to the surface area inside the transmission port.
6. Repeat the previous step for the reference beam at the reference beam entrance port.
7. If the reference and sample beam profiles meet the criteria specified in the previous
steps, the accessory optics are aligned properly and no further alignment is necessary.
Return to Step 16
of the Installation procedure. If the accessory alignment does not meet
the criteria, proceed to the following section.
Aligning the DRA
This procedure is used to align the transfer mirrors in the accessory optical chamber. A
one inch square of translucent paper is required that clearly displays the outline of the
sample or reference beam when illuminated. The following procedure should be
performed using white light operation. Refer to the illustration in the figure below when
performing the optics alignment.