Figure 37. The beam from the small spot kit with the iris partially open.
14. Remove the beam blocker from the reference beam path.
15. Replace the reflectance plug. The small spot kit is now aligned.
16. Turn off the instrument.
17. Plug in the DRA using the connector at the back of the sample compartment.
18. Fully open the iris and replace the accessory cover.
19. Turn on the instrument and allow it to initialize.
20. Return the iris to the position noted in step 13.
Powder cell
In some cases, a reflectance measurement may deliver better diffuse reflectance spectra
when the sample is ground to powder form. As shown in Figure 38 and 39, the powder
cell is designed to fit inside the reflectance sample holder in the same manner as the
reflectance standard. Each sample holder comes as a set: one cell with a Spectralon insert
for use as a reflectance standard, and an empty cell for loading the powdered sample. The
quartz window for each cell is compatible with each DRA accessory across the entire
250-2500 nm wavelength range.