Figure 49. The beam falling on the reference mounted in the jaw style centre mount.
4. Examine the path of the reference beam inside the integrating sphere. The beam
should strike the reference port without clipping any part of the sample or sample holder.
5. Remove the sample holder from the sphere and replace the sample with a calibrated
diffuse reflectance standard. If a calibrated standard is not available, use an uncalibrated
Spectralon standard. Load the sample holder assembly into the integrating sphere and
install the accessory cover.
6. Collect a Baseline scan
.
7. Mount the sample into the centre-mount sample holder. For the jaw style device, the
transmitted component from the sample beam must be eliminated so it is not collected by
the sphere. This can be accomplished by backing the sample with a piece of black felt
cloth. Load the sample holder assembly back into the sphere.
8. Click the Setup button and check the parameters. Normally, the setup parameters for
the application should match those used for the baseline scan. If using a previously
defined baseline correction, retrieve the correct baseline.
9. Click the Start button.
10. If the standard reference option was used for the baseline scan, the reflectance
spectra is generated automatically. If the simple or zero baseline correction was selected,
multiply the sample scan data by the spectral reflectance factors for the calibrated
reflectance standard or the typical data in Appendix B. Reflectance standards at
Labsphere are calibrated using 8°/h geometry