Options tab
SBW/Energy: Fixed
SBW
ON
SBW/Energy: SBW
(nm)
2.00
SBW/Energy: Beam
mode
Double
SBW/Energy: Slit
height
Reduced
Source: Lamps UV-Vis
Source: Source
changeover (nm)
350.0
Baseline tab
Correction
Zero
baseline/correction
Autostore tab
File Storage: Storage
Storage on
(prompt at start)
2. Place the specular port plug into position.
3. Place the appropriate PTFE reference plate over the sample port.
4. Select the Baseline button from the Scan dialog. Follow the on-screen prompts to
perform a 100%T baseline scan and a 0%T baseline scan.
Hot Tip
When performing specular measurements, block the sample beam in the
transmission position with a black masking.
5. Once the baseline correction is complete, the Ordinate status display in the top left
corner of the dialog will show ’Zero baseline’ in red text.
6. Clamp the sample over the sample port. The sample should be homogeneous,
substantially flat, and large enough to completely cover the image.
7. Press the Start button to perform the scan of the total (specular and diffuse)
reflectance of the sample.